6320 & 6340 SERIES PLANT GROWTH CHAMBERS OPERATIONS MANUAL FOR MODELS 6320, 6321, 6322, 6323 6340, 6341, 6342, 6343 PO Bo
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 10 of 61 SECTION 2 – EQUIPMENT OVERVIEW Congratulations! You have just pu
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 11 of 61 SECTION 2 – EQUIPMENT OVERVIEW – CON TINUED Humidity Control Panel (op
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 12 of 61 SECTION 3 – INSTALLATION Unpacking Your new unit has been thoroughly
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 13 of 61 Preliminary Cleaning Your new environmental chamber was thoroughly cl
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 14 of 61 Leveling the Unit Place a level on the middle shelf of the chamber. A
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 15 of 61 Connect each chamber to a grounded circuit. Failure to do so could res
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 16 of 61 Lights should only be used in a non-condensing environment. Connecting
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 17 of 61 SECTION 4 – ACCESSORY INSTALLATION Connecting Alarm Contacts (ALRM301)
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 18 of 61 Installing Carboy Water System (BOTL301) Humidified models can be purc
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 19 of 61 High concentrations of carbon dioxide can cause asphyxiation. The use
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 2 of 61 Dear Valued Customer: Thank you for purchasing CARON Products & Se
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 20 of 61 Connecting Communications or Analog Outputs (DLUX301 & DLUX302) Wi
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 21 of 61 6022, 6023, 6041, 6042 & 6043 have one controller each.
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 22 of 61 Use only distilled or deionized water with a resistivity between 50K-C
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 23 of 61 Installing Drain Water Pump (PUMP301) For humidified units in applic
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 24 of 61 SECTION 5 – OPERATION With the chamber properly installed and the appr
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 25 of 61 Controlling the Fluorescent & LED Lighting The plant growth chambe
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 26 of 61 Operating the timer The power switch enables and disables the lights
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 27 of 61 Examples for how to start your timer during day or night modes Night
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 28 of 61 Accumulate or Reset Factory default is set so the timer continu
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 29 of 61 Replacing Fluorescent Light Lamps The fluorescent light lamps have a m
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 3 of 61 TABLE OF CONTENTS Section 1 – Warranty ...
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 30 of 61 Section 6 – ACCESSORY OPERATION Using the Carboy Water System (BOTL301
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 31 of 61 Changing the CO2 Set-point (CO2C301) If an alternative CO2 set-point
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 32 of 61 Once the controllers are unlocked on each controller, follow these step
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 33 of 61 Analog Outputs: With DLUX301 & DLUX302 controllers, there is
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 34 of 61 Set the Scale Low value to correspond with the minimum value of th
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 35 of 61 Ramp & Soak: A ramp and soak control system is included wi
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 36 of 61 Changing the Humidity Set-point (HUMD301) To set the
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 37 of 61 Interior Electrical Outlet (OUTL301 & OUTL302) An optional interi
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 38 of 61 Operation of Front Mounted 6” Recorders (RCDR301, RCDR302) Built in
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 39 of 61 cartridge and metal pen arm be removed as an assembly. Unsnap the plast
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 4 of 61 Section 7 – Calibration ...
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 40 of 61 If a calibration adjustment is required for a single pen recorder, use
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 41 of 61 Operation of Front Mounted 10” Recorders (RCDR303, RCDR304) Built in
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 42 of 61 save all of the changes that have been made to the recorder&
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 43 of 61 Recorder Calibration: If calibration is required for single input re
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 44 of 61 Operation of Side Mounted Recorders (RCDR305, RCDR306) Side mounted H
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 45 of 61 3. Lift the chart from the hub and locating pin and slide it from unde
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 46 of 61 SECTION 7 – CALIBRATION The temperature and humidity systems can all b
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 47 of 61 Calibrating the Humidity If humidity calibration is needed, the follow
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 48 of 61 SECTION 8 – ALARMS Alarm System Overview The chamber control system i
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 49 of 61 SECTION 9 – PREVENTATIVE MAINTENANCE The CARON chamber has been robust
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 5 of 61 SECTION 1- WARRANTY INFORMATION EQUIPMENT LIMITED WARRANTY Please revi
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 50 of 61 SECTION 10 – SPECIFICATIONS MODEL 6320 6321 6322 6323 6340 6341 6342 6
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 51 of 61 SECTION 11 – ELECTRICAL SCHEMATICS
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 52 of 61 SECTION 11 – ELECTRICAL SCHEMATICS (CONTINUED)
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 53 of 61 SECTION 12 – TROUBLESHOOTING Problem -- Unit will not turn on Is the
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 54 of 61 SECTION 13 – SPARE / REPLACEMENT PARTS General Part Number Descriptio
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 55 of 61 Fuse Related . Lighting Related Part Number Description LGT-154 Bal
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 56 of 61 The temperature and humidity controllers are factory programmed for pre
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 57 of 61 APPENDIX A – RAMP & SOAK EXAMPLE General An efficient way to prog
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 58 of 61 8. Select the step type by pressing the ADVANCE button. 9. Use the
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 59 of 61 Event Output 1 Ent1 oFF Event Output 2 Ent2 oFF Step 2 P1 2 Soak ste
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 6 of 61 This writing is a final and complete integration of the agreement betwee
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 60 of 61 End step type S.tyP End Type of profile end (hold 30ºC) End HoLd Ste
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 61 of 61 Caron Products and Services, Inc. 27640 State Route 7 Marietta,
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 7 of 61 EQUIPMENT INTERNATIONAL LIMITED WARRANTY Please review this section bef
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 8 of 61 This writing is a final and complete integration of the agreement betwee
6320 & 6340 Series Operations Manual Rev G 9/16/2013 Page 9 of 61 INTERNATIONAL SYMBOLS AND DEFINITIONS WARNINGS